The Deckma OMD-11 Oil Content Monitor (OCM) is a critical piece of marine equipment designed to prevent ocean pollution. It continuously measures the oil concentration in bilge water before discharge.
Help you find a for an alarm code. Tell you where to find the full manual PDF . Explain the difference between the OMD-11 and OMD-2008 . Let me know how I can help! OMD-11 Oil-in-Water Monitor Manual | PDF | Relay - Scribd
: The unit uses plug-in technology for its optical components and circuit boards, making component replacement relatively straightforward.
Check that the desiccator crystal indicator is blue. If it is pink, it must be replaced. Ensure all piping connections are tight and free of leaks. Daily Operation and Flushing deckma omd11 manual
recommends using a dedicated cell cleaning brush and keeping spare desiccators and O-rings on hand. Calibration Check
This comprehensive guide serves as an operational reference based on the official protocols, detailing the device specifications, installation requirements, system programming, maintenance routines, and troubleshooting procedures. Key Technical Specifications
It should be mounted on a rigid vertical surface, ideally at eye level for easy monitoring. The Deckma OMD-11 Oil Content Monitor (OCM) is
Route the outlet pipe to an open funnel or back to the bilge holding tank.
Mount the OMD-11 as close to the oily water separator as possible, typically within 8 meters.
| Parameter | Specification | | :--- | :--- | | | Light Scattering (Infrared/Near-IR) | | Measuring Range | 0 – 100 ppm (parts per million) | | Resolution | 0.1 ppm | | Accuracy | ±10 ppm (0-100 ppm range) | | Sample Flow Rate | 3 – 4 Liters per minute (typical) | | Ambient Temperature | -25°C to +55°C | | Sample Water Temp | +10°C to +65°C | | Power Supply | 100-240 VAC, 50/60 Hz | | Alarm Set Points | 15 ppm (Standard Bilge), 100 ppm (Alarm) | | Outputs | 4-20 mA analog output; Relay contacts for alarms | | Protection Rating | IP56 / IP65 (Control Unit) | Tell you where to find the full manual PDF
The OMD-11 is equipped with to provide a margin of safety. The manual details the exact procedure for adjusting their parameters using the front panel buttons:
Do not open the cell cap while the system is hot; high temperatures can cause scalding.